FUNCTIONS
5.1 AUTO POWER OFF
The auto power off helps conserve power when using on battery or the AC adapter,
the machine will automatically turn off after 4 min of inactive use or no change in
the displayed weight.
6.0 CALIBRATION
Before calibrating the Dune series it is best to turn on the unit for about 1 min to let
the components warm up.
1) Press [on/off] key to turn the power on, if it is not already.
2) Press and hold the [Unit] key, until “CAL” is displayed
3) Remove any weight from the pan.
4) Press the [Unit] key, the display will flash “CAL” them will show the cal mass
required. If the cal mass is not displayed then the machine is not stable or is
not within a zero range. Turn off the unit and try zeroing the display before
trying another calibration.
5) Place the appropriate calibration mass on the scale